Power | - Voltage: Single-phase 208 - 240 VAC / 50/60Hz, (25 A breaker installed)
- 3 KW max.
- Note: Power cable included but without the plug. Please install a plug by your own or click the link below the picture to choose a plug according to your country or your lab requirements
|
Working Temperature | <= 1000°C continuously - Max. 1100°C, < 30 minutes
- Temperature Uniformity: +/- 2°C
- GE214 quartz tube is available for 1150oC working temperature at extra cost.
|
Heating Rate | - Max. 20°C/ min
- Recommended: 10°C/ min
|
Heating Elements | - High-quality Ni-Cr-Al resistance wire
|
Quartz Tube & Effective Heating | |
Vacuum Flange | - Stainless steel vacuum flange with dual high-temperature silicone O-rings with water cooling jacket
- One 1/4 gas inlet and the outlet port are built on the top of the flange with needle valves
- One bored Al2O3 foam refractory block (190Dia x 50mm) is provided
- K25 port is built-in for connecting to the vacuum pump
- Mechanical Vacuum Gauge (included as standard accessory), which ranges from -0.1 to 0.15Mpa
|
Water Cooling | |
Temperature Controller | - NRTL certified digital temperature controller is included
- Proportional–integral–derivative control (PID control) and auto-tune function
- 30 segments programmed with ramping, cooling, and dwelling steps
- Built-in over-temperature alarm and thermocouple failure alarm
- +/- 1 ºC temperature control accuracy
- THE Default DB9 PC communication port included for connecting to PC
- Eurotherm Temperature Controller is available at the extra cost with +/-0.1°C accuracy
|
Options | - Digital Pressure Gauge (optional) A gas-type independent(above 10mbar) digital vacuum gauge with a wide range between 3.8 x 10-5 to 1125 Torr. (Click Pic. 1 to order)
- You may choose a gas mixer to deliver multiple gasses with control over gas composition and flow rate (click pictures below to order separately).
Pic. 1Pic. 2 |
Vacuum Pressure | - Vacuum Level is dependent on the vacuum pump used, connected pipe, interior refractory block's materials, etc. By using KFD25, stainless steel pipe, MTI's two-stage mechanical pump, and KF25 Ball valve, the furnace can achieve vacuum levels below:
- 10-2 torr via mechanical pump within 30 minutes without thermal Block
- 10-4 torr via turbo-pump within 30 minutes with the Quartz Thermal Block at 900C
- Optional:
- The vacuum pump and pipe are not included, please click Pic.1 to order
- For achieving a higher vacuum level, a Turbopump Station is necessary. (click Pic.2)
- Spare sealing O-ring is available at extra cost (click Pic.3)
- KF25 Right-Angle Valve for quick connection to KF25 Bellows (click Pic.4)
Pic. 1 Pic. 2 Pic. 3 Pic. 4 |
Options | You may choose a gas mixer to deliver multiple gasses with control over gas composition and flow rate (click pictures below to order separately). |
Overall Dimensions | (click picture to see details) |
Net Weight | 80 kg |
Shipping Weight | 200 lbs |
Shipping Volume | 45"(L) x 28"(W) x 35"(H) |
Warranty
| - One-year limited warranty with lifetime support (consumable parts such as processing tubes, and O-rings are not covered by the warranty, please order replacements at related products below.).
- ATTENTION: Any damages caused by the use of corrosive and acidic gasses are not under the coverage of MTI One Year Limited Warranty.
|
Compliance | - CE Certified
- NRTL or CSA certification is available upon request at extra cost.
|
Warning Click here to learn the installation of a gas regulator
| - Tube furnaces with quartz tubes are designed for use under vacuum and at low-pressure < 0.2 bars
- Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation.
- A maximum temperature of 1000°C should be imposed on all quartz tube furnaces to avoid the quartz tubes undergoing deformations under vacuum.
- The flow rate for gasses should be limited to < 200 SCCM for reducing thermal shocks
|
Operation Demo Video | |