Furnace structure 
|
|
|
Power & Voltage
|
|
|
Max. Temperature
|
1200 oC
|
|
Continuous Temperature
|
1100 oC
|
|
Max. Heating Rate
|
<= 20 oC /min
|
|
Thermal Block Placement
|
 |
|
Heating Zone
|
-
Five independent heating zones:
-
Each heating zone length of 152 mm
-
Total heating zone length: 760 mm, ( nominal 800 mm L considering spacer )
- Max. Constant temperature zone 25" ( 635mm ) if 5 zone heated at same temperature ( within +/- 2oC)
|
Temperature Controllers 
|
- PID automatic control with 30 programmable segments for precise control of heating rate, cooling rate and dwell time.
- Built-in overheating & broken thermocouple protection.
- Over-temperature protection and alarm allows for operation without attendant(s).
- +/- 1 ºC temperature accuracy.
- Five K-type thermocouples (one for each zone).
- RS485 Communications Port.
- Optional:
- The furnace may be operated by PC by installing five control modules. Please order the MTS-02 control modules at extra cost.
- If you order the PC control module together with the furnace, we will install and test it before shipping. We recommend you to order a computer laptop from us to avoid software installation problems. For more information please click here.

|
Optional Processing Tube
& Vacuum flange
|
-
The furnace can accept quartz or mullite tube with diameter 1" - 4“ diameter and minimum length 1400mm ( 55"), which is not included in the package. Please click the picture below right to order tube & tube adapter, as well thermal blocks ( Note: To accept 5" Dia tube, customer needs to contact MTI tech. the support team for details)

|
Product Dimension  |
- Furnace:1035mm (L)x 385mm (W)x 370mm (H)
- Control box: 500mm (L)x 490mm (W)x 250mm (H)
|
| Net Weight |
About 85 Kg
|
| Warranty |
- One year limited warranty for electrical parts.
- (Consumable parts such as processing tubes, o-rings, and heating elements are not covered by the warranty, please order the replacement at related products below.)
|
| Laptop, software & WiFi Control(Optional) |
|
|
Compliance |
- CE Certified
- All electric components ( >24V) are UL / MET / CSA certified
- The furnace is ready to pass TUV(UL61010) or CSA certification at extra cost. (pls click marks below to learn details )
|
| Shipping Weight & Dimensions |
|
| Options (click picture to order at extra cost) |
|
| Notice
|
|
| Operation Instruction |
 |
|
Application Note
|
- The tube furnaces with quartz/mullite tube are designed for using under vacuum and low pressure <0.2 bars / 3 psi / 0.02 MPa
- Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation. Click here to learn the installation of a gas regulator.
- * Vacuum pressures may only be safely used up to 1000°C
- The flow rate for gasses should be limited to<200 SCCM (or 200ml/min) for reducing thermal shocks to the tub
|