Phone

(+65) 6284 3818

Email

sales@premier-sols.com

Enquriy

CS120 Potentiostat

0.0/5 rating (0 votes)

Description

Product Introduction
CS120 potentiostat contains a fast digital function generator, high-speed data acquisition circuitry, a potentiostat. With high performance in stability and accuracy with advanced hardware and well-functioned software, it is a research platform for corrosion, batteries, electrochemical analysis, sensor, life science and environmental chemistry etc.
Application
● Reaction mechanism of Electrosynthesis, electrodeposition, anodic oxidation, etc.
● Electrochemical analysis and sensor;  
● New energy materials (Li-ion battery, solar cell, fuel cell, supercapacitors), advanced functional materials, photoelectronic materials;
● Corrosion study of metals in water, concrete and soil, etc;
● Fast evaluation of corrosion inhibitor, water stabilizer, coating and cathodic protection efficiency. Standard supply list for each set
Instrument host CS120  x1
CS studio software x1
Power cable x1
USB cable x1
Cell cable x2
Dummy cell(1kΩ||100µF) x1
Manual x1 Service: (**all the service is free)
1. Warranty period: 5 years
2. Provide installation guidance and manual, software installation video.
3. Lifetime free software upgrading and technical service
4. Provide repair service for free Technical advantages
1. Polarization curve
Tafel plot can be obtained. The user can set the anodic reversal current (passivation film breakdown current) of the cyclic polarization curve to obtain material’s pitting potential and protection potential and uate the its susceptibility to intergranular corrosion. The software uses non-linear fitting to analyze polarization curve, and can make fast evaluation of material’s anti-corrosion ability and inhibitors.
Polarization curve of Ti-based amorphous alloy & stainless steel in 3%NaCl solution
2. Voltammetry
Linear Sweep Voltammetry(LSV), Cyclic Voltammetry(CV), SCV, SWV, DPV, NPV,ACV, Stripping voltammetry etc. It integrates calculation of peak area, peak current and standard curve analysis.
  
          
                                                                                                   
                                                                                                  LSV: mesoporous carbon material in 0.1M KOH                                  CV of PPy supercapacitor in 0.5 mol/L H2SO4
 
3. Full floating measurement
Full-floating mode be used for autoclave electrochemical measurements, on-line corrosion monitoring of metallic components under the ground (rebar in concrete, etc.) 4. Software development kit (SDK) 
  We are able to provide API functions and development examples, which facilitates some users’ requirements for secondary development and self-defined measurements. We can provide .dll file.
Software Features
Cyclic voltammetry: CS studio software provides users a versatile smoothing/differential/ integration kit, which can complete the calculation of peak height, peak area and peak potential of CV curves. In CV technique, during the data analysis, there is function of selecting exact cycle(s) to show. You can choose to see a cycle or some cycles as you want. You can also export data or vector graph of an exact cycle or several cycles.
Tafel plot and corrosion rate: CS studio also provides powerful non-linear fitting on Butler-Volmer equation of polarization curve. It can calculate Tafel slope, corrosion current density, limitation current, polarization resistance, corrosion rate. It can also calculate the power spectrum density, noise resistance and noise spectrum resistance based on the electrochemical noise measurements.
 
Real time saving of data: The data can be automatically saved even in case of sudden power off. Combination test: it can facilitate the automation of experiments and save time. With the unique function of combination test, you can choose several techniques, and set the wait time, the start time, and the cycles. Choose the experiments you want to run. Then you can make auto measurement of the set experiments as you want without having to wait in the lab. This function is especially useful if you have multi experiments to run and save your time greatly.
Data open: You can open the data files by .txt format in notepad. Data can also be opened in Origin
Specifications
Support 2-, 3- or 4-electrode system
Potential control range: ±10V
Current control range: ±2A
Potential control accuracy: 0.1%×full range±1mV
Current control accuracy: 0.1%×full range
Potential resolution: 10μV(>100Hz), 3μV (<10Hz)
Current sensitivity: 1 pA
Potentiostat rise time: <1μs (<10mA), <10μs(<2A)
Reference electrode input impedance: 1012Ω||20pF
Current range: 2nA~2A, 10 ranges
Compliance voltage: ±21V
Maximum current output: 2.0A
CV and LSV scan rate: 0.001mV~10,000V/s
Current increment during scan: 1mA@1A/ms<br< a=""> style="margin: 0px; padding: 0px;" />Potential increment during scan: </br<>0.076mV@1V/ms<br< a=""> style="margin: 0px; padding: 0px;" />AD data acquisition: </br<>16bit@1 MHz,20bit@1 kHz
DA Resolution: 16bit, setup time: 1μs
Minimum potential increment in CV: 0.075mV
Low-pass filters: covering 8-decade
Potential and current range: automatic
Communications Interface: isolated Universal Serial Bus (USB)
Operating System: Windows 2000/XP/ Win7/Win8/Win10
Techniques
Stable polarization
Open Circuit Potential (OCP), Potentiostatic (I-T measurement), Potentiodynamic (Tafel plot),
Transient polarization
Multi-Potential Steps, Potential Stair-Step (VSTEP)
Voltammetry
Cyclic Voltammetry (CV), Linear Sweep Voltammetry (LSV)

Share this product

Register for new product information:    

Premier Solutions (PS) Pte Ltd Singapore

Please publish modules in offcanvas position.