Two Heating Zone Tube Furnace | - Two heating zone with 440 mm length single heating zone and 250 mm length constant temperate zone
- 1200oC Max. working temperature for < 60 minutes
- 1100oC Max for continuous heating
- 30 segments programmable precision digital temperature controller with +/1ºC accuracy
- (Eurotherm 3000 Temperature controller is optional for +/- 0.1ºC accuracy)
- High purity quartz tube: 80mm OD x 72mm ID x 1400 mm Length 2.83"ID x 55'' Length
- 100 mm OD x 94mm ID tube is available as an option
- Input power: 208 - 240V AC input, a single phase at max. 3KW
- Three heating zone furnace is available as optional
|
Plasma RF Power Supply | - Output Power: 5 - 300W adjustable with ± 1% stability
- RF frequency: 13.56 MHz ±0.005% stability
- Matching: Automatic
- RF Output Port: 50 Ω, N-type, female
- Noise: <50 dB.
- Cooling: Air cooling
- Power: 208-240VAC, 50/60Hz, single phase
|
R2R Vacuum Chambers
| - Two vacuum chambers are located in two ends of the furnace and connected with the quartz tube
- The reeling rollers for a substrate in and out are located inside the vacuum chamber
- Reeling speed for the substrate is adjustable from 1- 400mm/min
- Max. movable substrate: 65 mm Width for 80mm OD tube
- One roll of copper foil with 65 mm width and 0.025 mm thick (~5 kg) is included for customer practice
- 100mm OD tube is available upon request, which can feeding substrate with max. width 80 mm
- Roll to Roll mechanism with speed controller and sealed flange without furnace may available upon request.
|
Three Channel Gas Delivery Station | - Working Temp.: 5~45 ºC
- Max. Pressure: 3x106 Pa
- Accuracy: ±1.5% FS
- Liner: ±(0.5-1.5)%F.S.
- Duplicate accuracy: ±0.2%F.S.
- Controlling Range:
- Controller 1: 0~100 SCCM (Ar gas calibration)
- Controller 2: 1~199 SCCM (H2 gas calibration)
- Controller 3: 1~499 SCCM (Methane calibration)
- Click Pic left for detail specification
- Warning: H2 and Methane must be installed nearby the furnace to alarm gas leaking.
|
Vacuum Pump and valve | - Heavy-Duty Rotary Vane Vacuum Pump (7.8 CFM -240 L/m) with Two-Stage Exhaust System installed in the bottom case with max. vacuum pressure 10-2 Torr.
- KFD25 adapter and stainless steel pipe are connected between the pump and tube flange with the ball valve
- Digital vacuum pressure gauge and display are installed with the furnace (Pic. 1)
- The oil-less pump (Pic. 2) and gas mixing & delivery system (Pic. 3) are available and PID pressure controlled vacuum system is available upon request at an extra cost
Pic. 1 Pic. 2Pic. 3 |
Anti-corrosive Pressure Gauge | - 3.8x10-5 to 1125 Torr measurement range
- Anti-corrosive, gas-type independent
- High accuracy and reproducibility at atmosphere for reliable atmospheric pressure detection
- Fast atmospheric detection eliminates waiting time and shortens process cycle
- Easy to exchange plug & play sensor element
- Click the picture to view detail spec
|
Temperature Controller | - Two temperature controllers with 30 programmable segments with +/- 1 ºC temperature accuracy
- Built-in PID Auto-Tune function with overheating & broken thermocouple protection
- Over-temperature protection and alarm allow for operation without attendant(s)
- RS485 Communications Port with PC operation software
- You may upgrade the temperature controller to the Eurotherm Controller with 0.1 ºC accuracy with software
|
Control Panel | - 7" touch screen control panel which will control temperature, gas flow rate, plasma power
- PC control software is available upon request for remote operation by Laptop computer (Must buy a laptop with pre-installed software)
|
Optional | - We may upgrade the furnace to 5" Tube, and grow 4" width film
- The larger diameter tube up to 8" OD for the roll to roll CVD system is available upon request
|
Overall dimensions | 2400 L x 600 W x 1250 H mm |
Warranty | One year limited warranty with lift time support (Consumable parts such as processing tubes, O-rings, and heating elements are not covered by the warranty, please order the replacement at related products below.) |
Compliance
| - CE Certified
- NRTL (UL61010) or CSA certification is available at extra cost.
|
Operation Instruction | |
Warning Click here to learn the installation of a gas regulator | - The quartz tube are designed for using under vacuum and low pressure < 0.2 bars / 3 psi / 0.02 Mpa
- Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation. Click here to learn the installation of a gas regulator.
- MTI reserves right to modify PECVD design at any time without notice but promise that quality will meet the specifications above
|
Total Shipping Weight | 500 kg (1100 lbs) including crate |