Phone

(+65) 6284 3818

Email

sales@premier-sols.com

Enquriy

UL Standard 1600C Three Zone Split Tube Furnace with Flange & 80 mm Dia. Alumina Tube - OTF-1600X-III-UL

0.0/5 rating (0 votes)

Description

OTF-1600X-III- is a split-able three-zone tube furnace with Kanthal 1650°C grade SiC heating element, 82mm OD Alumina tube and vacuum-sealed flange. Three 30 segments of temperature controllers are installed with S type thermal couples. Microprocessor-based self-tuning PID provides excellent control with no overshooting and  +/- 1°C accuracy. It is an excellent furnace for exchanging processing tube for multi-purpose up to 1600°C.

 

SPECIFICATIONS:

Power Input 9.0 K W
Working Voltage  AC220V Single-phase 50/60Hz ( 60 A breaker required)
Max. Working Temperature 1600°C  ≤1 hour
Continuously Working Temperature 1500°C
Recommend Heating rate 5°C/min.
Processing Tube

 Alumina Tube Size: 82 O.D x 73 I.D x 1000mm L 
 click the picture to order spare      click here to choose a crucible

Heat Zone Length
  • Three heating zones:  200 mm / each  ( included 20 mm spacing )
  • Total zone length:  600 mm  
  • Constant Temperature Zone length:   350mm  ( if set 3 zone at same temperature ) +/- 2ºC.
  • Accept processing tube up to 100 mm diameter
Temp. Profile at 1450ºC    (click to enlarge)
Overall Dimensions 860 mm × 380 mmW × 680 mmH
Temperature Controller

Temperature controller interface

The optional PC control module


Optional Eurotherm 3204 
Standard temperature controller:
  • Proportional–integral–derivative control (PID control) and auto-tune function
  • 30 segments programmed with ramping, cooling, and dwelling steps
  • Built-in over-temperature alarm and thermocouple failure alarm
  • Three S-type thermocouples.
  • +/- 1ºC temperature control accuracy
  • Default DB9 PC communication port (PC control module and software are sold separately. See optional.
Optional:  
European Temperature Controllers (optional upgrade)
  • 1 heating channel
  • 1 program of 8 or 24 segment
  • Alarm Relay: Form 'A' and Form ‘C’ built-in 
  • Digital Communications: RS485
  • Temperature Accuracy:  ± 0.1ºC
  • Recipes: Using a PC tool recipes can be created that can be used to change the operating parameters of the 3000 simply by selecting a new recipe via the HMI. This is very useful where multiple products are processed using the same controller but require different parameters to be set
Laptop, software & WiFi Control(Optional)
  • Brand new laptop with Microsoft Window 10 and Microsoft Office 2013 (30 days free trial) 
  • Labview Based Temperature Control System (EQ-MTS01) enables users to edit temperature profiles, manage heat-treatment recipes, record, and plot data for MTI furnaces
  • A wireless remote control provides up to 300 meters of operating range
  •     

Thermal Block Placement
  • Two thermal blacks are included. Please picture below to learn how to place them inside a tube
  • You may order flange support to avoid tube rotating (click Pic below right)
   
Vacuum Level
  • 10E-2 torr by a quality mechanical vacuum pump
  • 10-5 torr by molecular pump
Heating Elements & Thermal couples
  • 24 pcs Kanthal 1650ºC grade Silicon carbide rod (12 Dia. x 150 heating x  374 lengths, mm)
  • 3 pcs S type thermal couple built-in
  • Please click the picture below  to order a spare heating element
  • Notice: Customers will install SiC rod on-site by themselves as an instruction manual due to they are easy to damage during shipping
  •          
Power Cable & Plug
  • 10 feet length 10-3 AWG heavy duty UL approved power cable is included. Extra length cable can be ordered by click picture below-left
  • The plug is not included. Browse our catalog for compatible plugs or obtain one from a local hardware store. 
        
Net Weight 160 Kg
Operation Instructions
Warnings
  • The tube furnaces with alumina tube are designed for use under vacuum and low pressure < 0.02 Mpa 
  • Never use graphite crucible in an alumina tube.
  • Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation. Click here to learn the installation of a gas regulator.
  • The flow rate for gases should be limited to < 200 SCCM ( or 200ml/min )
  • Vacuum pressures may only be safely used up to 1500°C
Warranty
  • One year limited warranty with lifetime support (consumable parts such as processing tubes, and O-rings are not covered by the warranty, please order replacements at related products below.).  
  • ATTENTION: Any damages caused by the use of corrosive and acidic gases are not under the coverage of MTI One Year Limited Warranty.
Compliance
  • CE Certified
  • NRTL or CSA certification is available at the extra cost.

Share this product